Mikrobiyoloji Bulteni, vol.34, no.3-4, pp.215-221, 2000 (Scopus)
In this study, 102 staphylococci strains, 39 coagulase positive and 63 coagulase negative, were tested for oxacillin susceptibility by agar screening, microdilution, disk diffusion and automated Sceptor® system according to National Committee for Clinical Laboratory Standarts (NCCLS) recommendations. Tests were compared with each other by using agar screening as a reference method. Oxacillin resistance in staphylococci strains was found to be 60.8%, 57.8%, 58.8%, 54.9% in agar screening, microdilution, disk diffusion and automated Sceptor® system, respectively. The sensitivity, specificity and agreement rates of each test were found as follows when compared with the reference method; for microdilution 95.4%, 93.1%, 94.1%, for disk diffusion 95.3%, 95.2 %, 96.5 %, and for automated Sceptor® system 88.5 %, 88.8% and 87.2% respectively. In conclusion, to detect oxacillin resistance, sensitivity and specificity of automated Sceptor® system was found to be lower when compared to all others. Disk diffusion test can be used instead of agar screening test when it is not available.