Radar-Based Beat-Level Heartbeat Detection and Timing Validation Using Adaptive Matching


Aras S.

IEEE Access, cilt.14, ss.99123-99136, 2026 (SCI-Expanded, Scopus)

  • Yayın Türü: Makale / Tam Makale
  • Cilt numarası: 14
  • Basım Tarihi: 2026
  • Doi Numarası: 10.1109/access.2026.3708593
  • Dergi Adı: IEEE Access
  • Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus, Compendex, INSPEC, Directory of Open Access Journals
  • Sayfa Sayıları: ss.99123-99136
  • Anahtar Kelimeler: Adaptive matching, beat-level validation, beat-to-beat interval (IBI), cardiovascular monitoring, non-contact sensing, radar-based heartbeat detection, signal processing
  • Ondokuz Mayıs Üniversitesi Adresli: Evet

Özet

Systems capable of detecting micro-movements with high sensitivity, such as radar, allow for the non-contact measurement of vital physiological parameters, including respiratory and cardiac activity. However, low-amplitude displacements caused by heartbeats on the body surface are highly sensitive to noise. Therefore, reliable determination of heartbeats from radar signals remains a significant and current research challenge. This study presents a signal processing approach for determining the temporal positions of individual heartbeats from radar phase signals obtained through non-contact radar measurements. The proposed method involves improving heartbeat position estimation through peak detection and local maximum alignment based on envelope representation of the cardiac component, which is separated using phase unwrapping, differentiation, and band-pass filtering steps. The actual dataset we used includes a simultaneously recorded signal that can serve as a temporal reference for the heartbeat points obtained from the radar signal. Comparisons with the reference signal showed that the proposed method has a high level of detection accuracy with an average F1 score of 0.958 and an average timing error of 55 ms. Furthermore, the detected heartbeat intervals showed strong agreement with the reference signal, resulting in low error values. The results demonstrate that the proposed method offers competitive performance with methods in the literature for estimating average heart rate. However, the key contribution is the provision of temporal accuracy at the heartbeat level. This feature offers a significant advantage for analyses based on temporal intervals, such as HRV, and distinguishes it from similar studies in the literature.