Electroanalysis, vol.34, no.3, pp.474-484, 2022 (SCI-Expanded)
© 2021 Wiley-VCH GmbH.A novel solid-state electrochemical sensor based on a newly synthesized cross-linked copper(II) doped-copolymer and carbon nanotube material was developed for the direct determination of monohydrogen phosphate ions (HPO42−). The synthesized copolymers were characterized by FTIR, XPS, TG/DTG-DTA and SEM techniques. The sensor had a Nernstian slope:-30.7±0.4 mV/decade, linear concentrations range: 1.0×10−6 - 1.0×10−1 M, detection limit: 6.5×10−7 M, response time: 4 s and life time: 17 weeks. The sensor displayed constant potentials in the pH range 7.0-9.5. The sensor was successfully used as an indicator electrode in potentiometric titration and the direct determination of HPO42− in water samples.